INNER SHEATH, FOR 26 FR. OUTER SHEATH
Report
- Report Number
- 9610773-2023-02857
- Event Type
- Malfunction
- Date Received
- October 10, 2023
- Report Date
- October 23, 2023
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HIH
- UDI-DI
- 04042761029339
- PMA / PMN Number
- K931994
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. FURTHERMORE, MINOR SIGNS OF USE WERE NOTED ON THE INSTRUMENT. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE ROOT CAUSE OF THE EVENT IS LIKELY DUE TO PHYSICAL STRESS. HOWEVER, THE ROOT CAUSE OF THE REPORTED EVENT IS UNABLE TO BE DETERMINED. THE EVENT CAN BE DETECTED AND PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE (IFU) SECTIONS WHICH STATE: WARNING: INFECTION CONTROL RISK. PROPERLY REPROCESS THE PRODUCT BEFORE FIRST AND EACH SUBSEQUENT USE FOLLOWING THE INSTRUCTIONS IN THIS MANUAL AND IN THE SYSTEM GUIDE ENDOSCOPY. IMPROPER AND/OR INCOMPLETE REPROCESSING CAN CAUSE INFECTION OF THE PATIENT AND/OR MEDICAL PERSONNEL. 4.1 INSPECTION AND TESTING: INSPECTING THE PRODUCT. VISUALLY INSPECT THE PRODUCT. MAKE SURE THAT IT HAS: -- NO CORROSION. -- NO DENTS. -- NO SCRATCHES. CERAMIC INSULATION AT DISTAL END VISUALLY INSPECT THE CERAMIC INSULATION AT THE SHEATH¿S DISTAL END BEFORE EACH USE. DO NOT USE THE INSTRUMENT IN CASE OF DAMAGE (E.G. CRACKS, FRACTURES). WARNING. RISK OF INJURY. IMPACT, FALL, SHOCK OR SIMILAR STRESS CAN DAMAGE THE CERAMIC INSULATION AT THE SHEATH¿S DISTAL END. DAMAGED INSTRUMENTS CAN CAUSE INJURIES TO THE PATIENT AND/OR USER. DO NOT USE THE INSTRUMENT IF DAMAGED. DAMAGED PRODUCT: IF THE PRODUCT IS DAMAGED OR DOES NOT FUNCTION PROPERLY, CONTACT AN OLYMPUS REPRESENTATIVE OR AN AUTHORIZED SERVICE CENTER.¿ OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
G4 PMA/510(K) NUMBER: K931994/ K931995 TO DATE, THE DEVICE HAS NOT BEEN RETURNED TO OLYMPUS FOR EVALUATION. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.
THE CUSTOMER REPORTED TO OLYMPUS, THE INNER SHEATH CERAMIC TIP OF THE INNER SHAFT OF THE BACKFLUSH RESECTOSCOPE BROKE OFF. THERE WERE NO REPORTS OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332159 | INNER SHEATH, FOR 26 FR. OUTER SHEATH | - | HIH | OLYMPUS WINTER & IBE GMBH | A22040A | 23514 | 04042761029339 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |