TALENT THORACIC STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2010-01515
- Event Type
- Malfunction
- Date Received
- August 6, 2010
- Date of Event
- June 1, 2010
- Report Date
- July 9, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P070007
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): EVAL RESULTS: DISSECTION; SEVERELY CALCIFIED VESSELS. CONCLUSIONS: SEVERELY CALCIFIED VESSELS.
A TALENT THORACIC STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC ANEURYSM AND A TYPE 1 B DISSECTION WHICH STARTED BELOW THE THORACIC ARCH APPROXIMATELY 2 MONTHS AGO. THE EXTERNAL ILIAC AND COMMON FEMORAL ARTERIES WERE SEVERELY CALCIFIED, AND THE ILIAC VESSEL DIAMETER WAS 7-9 MM IN DIAMETER. THE PT HAD A CABG AND BILATERAL CAROTID BYPASS PROCEDURES DONE IN THE PAST. IT WAS REPORTED THAT A TALENT THORACIC CLINICAL TRIAL DEVICE WAS USED AS THE PROXIMAL MAIN, TWO TALENT THORACIC DISTAL MAIN STENT GRAFTS AND ONE TALENT THORACIC DISTAL EXTENSION. THERE WAS DIFFICULTY ADVANCING THE TALENT DEVICE THROUGH THE CALCIFIED AREA. THE DELIVERY SYSTEM COULD NOT BE ADVANCED AND KINKED. THE DELIVERY SYSTEM WAS REMOVED FROM THE PT. THE ARTERIOGRAM SHOWED A SHORT SEGMENT DISSECTION IN THE RIGHT EXTERNAL ILIAC ARTERY BELOW THE INTERNAL AND COMMON ARTERY BIFURCATION. A 9 X 40 MM BRIDGE ASSURANT STENT WAS PLACED OVER THIS DISSECTED AREA WITH EXCELLENT RESULTS, AND THERE WAS GOOD FLOW. UPON REMOVAL OF ALL THE SHEATHS AND INTRODUCERS FROM THE RIGHT SIDE, THERE WAS A FOCAL DISSECTION IN WHERE THE STENT GRAFT DELIVERY SYSTEMS WERE INSERTED (SEE MFR # 2953200-2010-01516, MFR # 2953200-2010-01517 NAD MFR # 2953200-2010-01518). THIS AREA IN THE COMMON FEMORAL ARTERY WAS EXCISED AND REPLACED WITH A GORE-TEX INTERPOSITION AND RESOLVED THE DISSECTION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT THORACIC STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR | NA | V00266809 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention |