FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 17896960 · Received October 9, 2023

Report

Report Number
1645337-2023-11948
Event Type
Injury
Date Received
October 9, 2023
Date of Event
September 30, 2022
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000334
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2023, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT CORRECT DATE OF IMPLANTATION WAS ON (B)(6) 2014. SECTION D 6A. IMPLANTATION DATE: (B)(6) 2014.

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON OCTOBER 17, 2023, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT WAS INITIALLY DIAGNOSED WITH LEFT BREAST ASYMMETRY, THEN RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE II), BILATERAL PTOSIS, BILATERAL RIPPLING, AND LEFT BREAST PAIN. ON OCTOBER 30, 2023, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON OCTOBER 31, 2023, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SM MPP GEL 450CC BREAST IMPLANT WAS FOUND TO BE RUPTURED. ADDITIONALLY, SHELL ABRASION WAS OBSERVED ON THE EDGES OF THE RUPTURE. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE WAS CONSISTENT WITH NORMAL WEAR. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE. THIS MAY BE THE RESULT OF THE FOLLOWING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE SUCH AS TOO SMALL BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IN SOME CASES, THE BREAST IMPLANTS MAY ALSO WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULES, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. CAPA ACTIVITY WAS NOT REQUIRED. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS. COMPLICATIONS ON THE LEFT BREAST WILL BE REPORTED UNDER MRN: 1645337-2023-13179.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 450CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT SUFFERED SPONTANEOUS RIGHT BREAST IMPLANT RUPTURE. AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT ON (B)(6) 2023. THE REPLACEMENT DEVICES WERE: (RIGHT) 685CC MENTOR MEMORYGEL BOOST BREAST IMPLANT CATALOG: SHPB685 LOT: 9933425 SN: (B)(6) AND (LEFT) 685CC MENTOR MEMORYGEL BOOST BREAST IMPLANT CATALOG: SHPB685 LOT: 9933425 SN: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
867461 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 6721258 00081317000334

Patients

Seq Age Sex Outcome Treatment
1 44 YR Female Required Intervention