FDA Adverse Event Malfunction Summary report: N

LP10 VOLUME VENTILATOR

MDR report key: 178953 · Received July 22, 1998

Report

Report Number
2183157-1998-00116
Event Type
Malfunction
Date Received
July 22, 1998
Date of Event
June 25, 1998
Report Date
July 22, 1998
Manufacturer
NELLCOR PURITAN BENNETT
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING SERVICE OF PRODUCT UNIT WAS FOUND TO NOT CYCLE WITH ALL LEDS AND CONSTANT SINGLE TONE AUDIBLE ALARM DUE TO INTEGRATED CIRCUIT U28 ON THE LOGIC BOARD BEING OUT OF SPECIFICATION. REPLACED U28.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LP10 VOLUME VENTILATOR VOLUME VENTILATOR CBK NELLCOR PURITAN BENNETT LP10 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other