ENDOPATH** XCEL
Report
- Report Number
- 3005075853-2010-04513
- Event Type
- Malfunction
- Date Received
- August 9, 2010
- Date of Event
- July 13, 2010
- Report Date
- July 13, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE B (B)(4), MFG DATE: 12/17/2009, EXP DATE: 01/17/2014. DEVICE C (B)(4), MFG DATE: 03/09/2010, EXP DATE: 02/09/2014. DEVICE D BATCH # UNK. DEVICE E (B)(4), MFG DATE: 12/17/2009, EXP DATE: 11/17/1014. DEVICE H (B)(4), MFG DATE: 03/09/2010, EXP DATE: 02/09/2014. DEVICE F (B)(4), MFG DATE: 03/09/2010, EXP DATE: 02/09/2014. LEAK TEST FAILED INSERTING AND REMOVING TEST PROBE. DEVICE G (B)(4), MFG DATE: 12/17/2009, EXP DATE: 11/17/1014. LEAK TEST FAILED WITHOUT TEST PROBE, DAMAGED DUCKBILL. ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION, THE DEVICE WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE ANALYSIS RESULTS FOUND THAT DEVICE (C) WAS RETURNED IN GOOD VISUAL CONDITION, THE DEVICE WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE ANALYSIS RESULTS FOUND THAT DEVICE (D) WAS RETURNED IN GOOD VISUAL CONDITION, THE DEVICE WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE ANALYSIS RESULTS FOUND THAT DEVICE (E) WAS RETURNED IN GOOD VISUAL CONDITION, THE DEVICE WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE ANALYSIS RESULTS OF DEVICE (F) FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. ONE POSSIBLE CAUSE FOR THIS TYPE OF ISSUE IS EXCESSIVE BENDING LOAD AS A RESULTING FROM SURGEON MANIPULATION OF INSERTED DEVICES. IN ADDITION, THE LENS OF THE OBTURATOR WAS NOTED CRACKED. THE EXPOSURE TO ETHYLENE OXIDE (EO) WHICH IS PART OF THE COMPLAINT DEVICE RETURN PROCESS MAY LEAD TO CRACK/SCRATCH LENS. THE ANALYSIS RESULTS FOUND THAT DEVICE (G) WAS RECEIVED WITH THE DUCKBILL SEAL OPEN AT THE SLIT. AS A RESULT, IT WOULD NOT OPEN AND CLOSE AS INTENDED DURING FUNCTIONAL TESTING. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. THE ANALYSIS RESULTS FOUND THAT DEVICE (H) WAS RETURNED IN GOOD VISUAL CONDITION, THE DEVICE WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE, ALL SIX DEVICES WERE LEAKING THROUGHOUT THE CASE. TWO DEVICES WERE PLACED AND WERE LEAKING THEN TWO MORE DEVICES WERE PLACED AND THEY LEAKED AS WELL. TWO MORE DEVICES WERE PLACED WITH ADDITIONAL DEVICES USED TO COMPLETE THE PROCEDURE WITH THE PORT SITES LEAKING THE ENTIRE CASE. ADDITIONAL PORTS WERE CREATED WITH NO CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** XCEL | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | G4RN30 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |