INPEN MMT-105NNBLNA NOVO NORDISK BLUE
Report
- Report Number
- 3012822846-2023-01667
- Event Type
- Malfunction
- Date Received
- October 9, 2023
- Date of Event
- September 13, 2023
- Report Date
- October 9, 2023
- Manufacturer
- COMPANION MEDICAL INC
- Product Code
- FMF
- UDI-DI
- 000010862088000344
- PMA / PMN Number
- K201337
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- 003
Narratives
PER VISUAL INSPECTION: NO PHYSICAL DAMAGE TO CARTRIDGE HOLDER OR INPEN FRONT AND BACK SHELL WAS NOTED. CARTRIDGE HOLDER LOCKS PROPERLY IN PLACE. INPEN PAIRED WITH COMMERCIAL MOBILE APP WITH SMALL DOSE. PERFORMED LEADSCREW RESET TORQUE TEST. INPEN WITHIN SPECIFICATION (CW: 0.4 OZF-IN AND CCW: 4.90 OZF-IN). PERFORMED ENCODER BOND INVESTIGATION AND FOUND THAT THE ENCODER PATTERN WHEEL TABS ROTATING AND TRAVELING OFF THE KEYED SLOTS OF DOSE NUT GUIDES. ENCODER PATTERN WHEEL SHOULD NEVER ROTATE. THIS CAUSES AN UNEXPECTED TRAVEL OF THE ENCODER PATTERN WHEEL CREATING RESISTANCE TO DIAL AND DISPENSING. ALSO, PLASTIC SHAVINGS CONTAMINATION BUILDS UP FOUND CAUSED BY ENCODER WHEEL TABS RUBBING AT THE WALLS OF THE DOSE NUT. ABRASIONS DOWN THE LENGTH OF THE DOSE NUT. UNABLE TO PERFORM BASELINE/WIRELESS FUNCTIONALITY AND DISPLACEMENT DOSE ACCURACY TEST. INPEN PASSED FRONT CAP INVESTIGATION. IN CONCLUSION: IT WAS DETERMINED FROM DESTRUCTIVE ANALYSIS THAT THE DIFFICULT TO DIAL/DOSE WAS CAUSED BY A PATTERN WHEEL MISALIGNMENT DUE TO AN ENCODER BASE BOND FAILURE. THIS CAN AFFECT INSULIN DELIVERY. THEREFORE, THE CUSTOMER CONCERN OF DOSE DIAL BEING DIFFICULT TO DIAL/DOSE WAS CONFIRMED. LEADSCREW ANOMALY WAS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THE SCREW MOVEMENT WAS NOT GIVEN DOSE PROPERLY. TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS NOT RESOLVED AND FOUND THAT RETRACTING THE SCREW WAS DIFFICULT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE DEVICE AND INPEN WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1121235 | INPEN MMT-105NNBLNA NOVO NORDISK BLUE | SYRINGE, PISTON | FMF | COMPANION MEDICAL INC | MMT-105NNBLNA | B0295 | 000010862088000344 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Female |