FDA Adverse Event Malfunction Summary report: N

PROPONENT MRI DR

MDR report key: 17886971 · Received October 6, 2023

Report

Report Number
2124215-2023-55324
Event Type
Malfunction
Date Received
October 6, 2023
Date of Event
September 30, 2023
Report Date
October 26, 2023
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LWP
PMA / PMN Number
P150012/S000
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS PATIENT'S REMOTE COMMUNICATOR DISPLAYED A RED CALL DOCTOR ICON. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) ASKED THE PATIENT TO PERFORM A PATIENT-INITIATED INTERROGATION (PII) AND AFTER TROUBLESHOOTING, A SUCCESSFUL PII WAS SENT. FURTHER INVESTIGATION FOUND THAT AN ALERT FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE LOW OUT OF RANGE HAD BEEN STORED. THE PATIENT WAS REFERRED TO THE CLINIC FOR EVALUATION AND NO ADVERSE EFFECTS WERE REPORTED. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. THE DEVICE REMAINS IN SERVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS PATIENT'S REMOTE COMMUNICATOR DISPLAYED A RED CALL DOCTOR ICON. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) ASKED THE PATIENT TO PERFORM A PATIENT-INITIATED INTERROGATION (PII) AND AFTER TROUBLESHOOTING, A SUCCESSFUL PII WAS SENT. FURTHER INVESTIGATION FOUND THAT AN ALERT FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE LOW OUT OF RANGE HAD BEEN STORED. THE PATIENT WAS REFERRED TO THE CLINIC FOR EVALUATION AND NO ADVERSE EFFECTS WERE REPORTED. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. THE DEVICE REMAINS IN SERVICE. ADDITIONAL INFORMATION WAS RECEIVED. THE PATIENT WAS CHECKED, BUT LEAD REVISION WAS NOT PERFORMED PER PHYSICIANS DECISION. LOW PACING IMPEDANCE, HIGH PACING THRESHOLD MEASUREMENTS AND NOISY SIGNALS, ARE KNOWN RV LEAD OBSERVATIONS FOR THIS PATIENT. THE DEVICE WAS REPROGRAMMED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE REMAINS IN SERVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS PATIENT'S REMOTE COMMUNICATOR DISPLAYED A RED CALL DOCTOR ICON. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) ASKED THE PATIENT TO PERFORM A PATIENT-INITIATED INTERROGATION (PII) AND AFTER TROUBLESHOOTING, A SUCCESSFUL PII WAS SENT. FURTHER INVESTIGATION FOUND THAT AN ALERT FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE LOW OUT OF RANGE HAD BEEN STORED. THE PATIENT WAS REFERRED TO THE CLINIC FOR EVALUATION AND NO ADVERSE EFFECTS WERE REPORTED. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE. THE DEVICE REMAINS IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
219140 PROPONENT MRI DR PACEMAKER LWP BOSTON SCIENTIFIC CORPORATION L211 708505

Patients

Seq Age Sex Outcome Treatment
1 Unknown