INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2023-16144
- Event Type
- Injury
- Date Received
- October 6, 2023
- Date of Event
- September 13, 2023
- Report Date
- November 27, 2023
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628029934
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: ¿ RUPTURE: OBSERVED THREE OPENINGS ON ANTERIOR ASSESSED AS SURGICAL DAMAGE. ADDITIONAL OBSERVATIONS: ¿ CREASE FLAT OBSERVED ON THE DEVICE.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. HEALTHCARE PROFESSIONAL REPORTED "HOLE, NON-SPECIFIC" VIA RGA. DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. HEALTHCARE PROFESSIONAL REPORTED "HOLE, NON-SPECIFIC" VIA RGA. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1270125 | INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 3551041 | 10888628029934 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Female | Required Intervention |