FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 17886057 · Received October 6, 2023

Report

Report Number
9617229-2023-16144
Event Type
Injury
Date Received
October 6, 2023
Date of Event
September 13, 2023
Report Date
November 27, 2023
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628029934
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: ¿ RUPTURE: OBSERVED THREE OPENINGS ON ANTERIOR ASSESSED AS SURGICAL DAMAGE. ADDITIONAL OBSERVATIONS: ¿ CREASE FLAT OBSERVED ON THE DEVICE.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. HEALTHCARE PROFESSIONAL REPORTED "HOLE, NON-SPECIFIC" VIA RGA. DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. HEALTHCARE PROFESSIONAL REPORTED "HOLE, NON-SPECIFIC" VIA RGA. DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1270125 INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3551041 10888628029934

Patients

Seq Age Sex Outcome Treatment
1 55 YR Female Required Intervention