INPEN MMT-105NNPKNA NOVO NORDISK PINK
Report
- Report Number
- 3012822846-2023-01656
- Event Type
- Malfunction
- Date Received
- October 6, 2023
- Date of Event
- September 9, 2023
- Report Date
- November 19, 2023
- Manufacturer
- COMPANION MEDICAL INC
- Product Code
- FMF
- UDI-DI
- 000010862088000368
- PMA / PMN Number
- K201337
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
PER VISUAL INSPECTION: NO PHYSICAL DAMAGE TO CARTRIDGE HOLDER OR INPEN FRONT AND BACK SHELL WAS NOTED. CARTRIDGE HOLDER LOCKS PROPERLY IN PLACE. THE INPEN PAIRED TO THE COMMERCIAL APP. INPEN PASSED BASELINE AND WIRELESS FUNCTIONALITY. APP LOGBOOK DISPLAYED: 15U, 15U, 15U, 15U, 15U, 15U, 15U, 15U, 15U, 15U. INPEN PASSED LEADSCREW RESET TORQUE TEST AND IS WITHIN SPECIFICATION (CW: 0.45 OZF-IN AND 4.00 OZF-IN). INPEN PASSED FRONT CAP INVESTIGATION. PENDING FURTHER INVESTIGATION PERFORMED IN SAN DIEGO LOCATION. IN CONCLUSION: PER SAN DIEGO LOCATION: INPEN PASSED BASE LINE FUNCTIONALITY TEST AND DISPLACEMENT DOSE ACCURACY. LEADSCREW RESET TORQUE (CW 2.37 AND CCW 2.35). PAIRED TO COMMERCIAL APP USING 4 UNITS. INPEN BASELINE AND WIRELESS FUNCTIONALITY. APP LOGBOOK DISPLAYED: 15,15,15,15,15,15,15,15,15,15. KNOB WAS CUT TO REVEAL ELECTRONICS HOUSING. NO SIGNS OF ABRASION WERE FOUND. ENCODER BASE BOND WAS STILL INTACT. NO PROBLEMS FOUND WITH THIS INPEN, ALL FUNCTIONS TESTED OK. LEADSCREW ANOMALY WAS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT THE INPEN WAS NOT DISPENSING ANY INSULIN. IT WAS STATED THAT THE SCREW WAS GETTING STUCK AND THE SCREW IN THE INPEN DID NOT MOVE WHEN THE BUTTON WAS PUSHED. TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE INPEN AND WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202239 | INPEN MMT-105NNPKNA NOVO NORDISK PINK | SYRINGE, PISTON | FMF | COMPANION MEDICAL INC | MMT-105NNPKNA | B1080 | 000010862088000368 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Female |