RAD-G
Report
- Report Number
- 3019388613-2023-00253
- Event Type
- Malfunction
- Date Received
- October 5, 2023
- Date of Event
- June 9, 2023
- Report Date
- September 12, 2023
- Manufacturer
- MASIMO - 15750 ALTON PKWY
- Product Code
- DQA
- UDI-DI
- 00843997000666
- PMA / PMN Number
- K201770
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL MANUFACURING NARRATIVE: OTHER, OTHER TEXT: THE RETURNED RAD-G WAS EVALUATED. THE ISSUE WAS CAUSED BY SHORTING INSIDE THE ON/OFF POWER BUTTON WHICH CREATED AN ELECTRICAL SHORT ACROSS THE BUTTON, AND RESULTED IN THE BUTTON BEING ACTIVATED EVEN WHEN NOT PHYSICALLY PRESSED.
MASIMO'S INITIAL REPORT IDENTIFIED THE MODEL AS NUMBER 27977. MODEL NUMBER 9849 IS ASSOCIATED WITH THE GUDID NUMBER. PART NUMBER 27977 IS THE SUBASSEMBLY TO PART NUMBER 9849. THIS SUPPLEMENTAL MDR UPDATES THE MODEL NUMBER TO 9849 AND BRAND NAME TO RAD-G TO ENSURE CORRELATION WITH THE GUDID DATABASE.
ADDITIONAL MANUFACURING NARRATIVE: THE PRODUCT HAS BEEN RETURNED TO MASIMO FOR EVALUATION. WHEN THE INVESTIGATION IS COMPLETE A FOLLOW UP REPORT WILL BE SUBMITTED.
THIS CORRECTION UPDATES THE UDI NUMBER TO INCLUDE THE DI AND PI FIELDS, ALL NUMBERS, LETTERS, PARENTHESES, AND SYMBOLS.
THE CUSTOMER REPORTED THE RAD-G TURNS OFF AND ON MULTIPLE TIMES. THERE WERE NO PATIENT IMPACT OR CONSEQUENCES REPORTED.
THE CUSTOMER REPORTED THEIR RAD-G DEVICE TURNS OFF AND ON MULTIPLE TIMES. THERE WAS NO PATIENT IMPACT OR CONSEQUENCES REPORTED.
THE CUSTOMER REPORTED THEIR RAD-G DEVICE TURNS OFF AND ON MULTIPLE TIMES. THERE WERE NO PATIENT IMPACT OR CONSEQUENCES REPORTED.
THE CUSTOMER REPORTED THEIR RAD-G DEVICE TURNS OFF AND ON MULTIPLE TIMES. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1342153 | RAD-G | OXIMETER | DQA | MASIMO - 15750 ALTON PKWY | 9849 | 00843997000666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |