FDA Adverse Event Malfunction Summary report: N

QUICKIE

MDR report key: 17882157 · Received October 5, 2023

Report

Report Number
9616084-2023-00058
Event Type
Malfunction
Date Received
October 5, 2023
Date of Event
September 6, 2023
Report Date
January 31, 2025
Manufacturer
SUNRISE MEDICAL (US) LLC
Product Code
IOR
PMA / PMN Number
K072153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

BACKGROUND INFORMATION: QUICKIE QX/QXI WHEELCHAIR OWNER'S MANUAL, REV. H, PAGE 14 STATES: "INSPECT WHEEL LOCKS WEEKLY PER THE MAINTENANCE CHART. DO NOT USE YOUR CHAIR UNLESS YOU ARE SURE BOTH WHEEL LOCKS CAN FULLY ENGAGE. A WHEEL LOCK THAT IS NOT CORRECTLY ADJUSTED MAY ALLOW YOUR CHAIR TO ROLL OR TURN UNEXPECTEDLY. WHEEL LOCKS MUST BE ADJUSTED AFTER MAKING SURE THE TIRES HAVE THE CORRECT AIR PRESSURE. WHEN FULLY ENGAGED, THE ARM SHOULD BE EMBEDDED INTO THE TIRE AT LEAST 1/8' TO BE EFFECTIVE" AND "IF YOU FIND THE WHEEL LOCKS HAVE SLIPPED OR ARE NOT WORKING CORRECTLY CONTACT YOUR SUNRISE MEDICAL AUTHORIZED DEALER FOR PROPER ADJUSTMENT." QUICKIE QX/QXI WHEELCHAIR OWNER'S MANUAL, REV. H, PAGE 28 STATES: "WE WARRANT ALL SUNRISE-MADE PARTS AND COMPONENTS OF THIS WHEELCHAIR AGAINST DEFECTS IN MATERIALS AND WORKMANSHIP FOR ONE YEAR FROM THE DATE OF FIRST CONSUMER PURCHASE." DISCUSSION: AFTER EVALUATING THE COMPLAINT, THE DEALER REPORTS THAT THE LEFT SIDE PUSH WHEEL LOCK HARDWARE IS BECOMING LOOSE. IT WAS DETERMINED THAT BASED ON SIMILAR COMPLAINTS AND PRODUCTS RECEIVED, THE POTENTIAL CAUSE COULD BE HARDWARE LOOSENING OVER TIME LEADING TO INSUFFICIENT WHEEL LOCKING FORCE. THIS POTENTIAL CAUSE IS CURRENTLY UNDER REVIEW AND ADDITIONAL INFORMATION ABOUT THE WHEEL LOCKS ARE BEING REQUESTED FOR REVIEW. BASED ON THE OWNER'S MANUAL, THE USER OR DEALER SHOULD BE PERFORMING MAINTENANCE TO REDUCE THE RISK OF WHEEL LOCKS BECOMING NON-FUNCTIONAL OR BROKEN. THE DEALER HAS RETIGHTENED THE HARDWARE MULTIPLE TIMES, BUT IT CONTINUES TO LOOSEN. THE AGE OF THE CHAIR AT THE TIME OF THIS COMPLAINT WAS APPROXIMATELY 126 DAYS. ACCORDING TO THE QUICKIE QX/QXI WHEELCHAIR OWNER'S MANUAL, SUNRISE MEDICAL PROVIDES A ONE-YEAR WARRANTY FROM DATE OF PURCHASE FOR ALL SUNRISE-MADE PARTS AND COMPONENTS THAT HAVE DEFECTS IN THE MATERIAL OR WORKMANSHIP. THE DEALER REQUESTED A REPLACEMENT OF THE LEFT-HAND WHEEL LOCK. THERE WERE NO INJURIES OR ADVERSE IMPACT TO THE USER REPORTED. CONCLUSION: IN CONCLUSION, THE LOOSENING OF THE WHEEL LOCK HARDWARE OVER TIME IS THE PRIMARY POTENTIAL CAUSE IDENTIFIED. THE INVESTIGATION INTO THE MATTER IS ONGOING. THE PRODUCT IN QUESTION MET ALL PRODUCT SPECIFICATIONS BEFORE RELEASE FOR DISTRIBUTION AT THE TIME OF SHIPPING TO THE CUSTOMER. THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THERE IS NO CLAIM OF INJURY IN THIS CASE. BECAUSE THE FAILURE MODE OF NON-FUNCTIONING WHEEL LOCKS HAS BEEN PREVIOUSLY REPORTED, PER 21 CFR 803.50, THIS MDR IS BEING FILED.

Additional Manufacturer Narrative · 0

UDI RELATED DATA QUALITY UPDATES ONLY. THIS MDR ONLY CONTAINED THE DI PORTION OF THE UDI INCLUDED IN THE MEDWATCH FORM. THE MDR HAS BEEN CORRECTED BY ADDING THE FULL UDI NUMBER.

Description of Event or Problem · 0

DEALER REPORTED THAT THE LEFT SIDE PUSH WHEEL LOCK HARDWARE IS BECOMING LOOSE. THE DEALER HAS RETIGHTENED THE HARDWARE MULTIPLE TIMES BUT IT CONTINUES TO LOOSEN. THE DEALER HAS REQUESTED A REPLACEMENT OF THE LEFT SIDE WHEEL LOCK. NO INJURIES OR ADVERSE IMPACT TO THE USER WAS REPORTED.

Description of Event or Problem · 0

THIS REPORT REPRESENTS A CORRECTION TO A PREVIOUSLY SUBMITTED MDR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2156903 QUICKIE MECHANICAL WHEELCHAIR IOR SUNRISE MEDICAL (US) LLC QUICKIE QXI

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown