FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 17875434
·
Received October 5, 2023
Report
- Report Number
- 3013756811-2023-137166
- Event Type
- Malfunction
- Date Received
- October 5, 2023
- Date of Event
- September 14, 2023
- Report Date
- October 30, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
THE FAILURE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE ANALYSIS, THE ALLEGED CARTRIDGE ALARM ISSUE WAS VERIFIED, BUT THE INSULIN GAUGE ISSUE COULD NOT BE VERIFIED.
Additional Manufacturer Narrative · 0
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. REPORTEDLY, CUSTOMER LOADED A NEW CARTRIDGE TO ADDRESS THE ISSUE. ADDITIONALLY, IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING INSULIN DELIVERY. CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 193-197 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 352527 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 60410363 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Female |