FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH¿ NORMAL SALINE SYRINGES,

MDR report key: 17873723 · Received October 4, 2023

Report

Report Number
1911916-2023-00735
Event Type
Malfunction
Date Received
October 4, 2023
Date of Event
September 20, 2023
Report Date
October 24, 2023
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
NGT
UDI-DI
30382903065463
PMA / PMN Number
K003553
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE AVAILABLE FOR EVAL?: YES. D.9. RETURNED TO MANUFACTURER ON: 19-OCT-2023 . H.6. INVESTIGATION SUMMARY: IT WAS REPORTED BLOOD BYPASSED THE RUBBER PLUNGER. TO AID IN THE INVESTIGATION, ONE EMPTY SAMPLE WAS RECEIVED WITH NO PACKAGING FLOW WRAP OR TIP CAP. A VISUAL INSPECTION WAS PERFORMED, AND NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE SAMPLE WAS THEN TESTED FOR LEAKAGE PAST THE STOPPER AND PASSED. IT COULD BE POSSIBLE THE CUSTOMER IS NOT USING THE UNIT AS INTENDED. THIS PRODUCT IS DESIGNED TO EXPEL THE SOLUTION BY PRESSING THE PLUNGER ROD DOWN, THEN DISPOSE OF THE SYRINGE. IT IS NOT DESIGNED TO PULL THE PLUNGER ROD BACK. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306546, LOT 3142731. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD POSIFLUSH¿ NORMAL SALINE SYRINGES, LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INCIDENT DETAILS: WAS FLUSHING PATIENT'S LUMEN AND BLOOD BYPASSED THE RUBBER PLUNGER AND SPRAYED BACK AT ME. IMPACT OF INCIDENT: NO HARM TO PATIENT. UNEXPECTED OR PROLONGED CARE? NO.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD POSIFLUSH¿ NORMAL SALINE SYRINGES, LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INCIDENT DETAILS: WAS FLUSHING PATIENT'S LUMEN AND BLOOD BYPASSED THE RUBBER PLUNGER AND SPRAYED BACK AT ME. IMPACT OF INCIDENT: NO HARM TO PATIENT. UNEXPECTED OR PROLONGED CARE? NO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
342692 BD POSIFLUSH¿ NORMAL SALINE SYRINGES, SALINE, VASCULAR ACCESS FLUSH NGT BD MEDICAL (BD WEST) MEDICAL SURGICAL 3142731 30382903065463

Patients

Seq Age Sex Outcome Treatment
1 Unknown