FDA Adverse Event Injury Summary report: N

ANATOMICAL SHOULDER SYSTEM METAL HUMERAL HEAD Ø46

MDR report key: 17868939 · Received October 4, 2023

Report

Report Number
3005180920-2023-00781
Event Type
Injury
Date Received
October 4, 2023
Date of Event
September 11, 2023
Report Date
October 4, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
KWS
UDI-DI
07630040713160
PMA / PMN Number
K170910
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 14 SEPTEMBER 2023: LOT 1903164: 20 ITEMS MANUFACTURED AND RELEASED ON 30-JUL-2019. EXPIRATION DATE: 2024-07-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 18 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICES INVOLVED BATCH REVIEW PERFORMED ON 14 SEPTEMBER 2023: ANATOMICAL SHOULDER SYSTEM 04.01.0004 STD HUMERAL DIAPHYSIS - CEMENTLESS - 9 (K170452) LOT 2012131: 26 ITEMS MANUFACTURED AND RELEASED ON 01-FEB-2021. EXPIRATION DATE: 2026-01-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 25 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ANATOMICAL SHOULDER SYSTEM 04.01.0026 HUMERAL ANATOMICAL METAPHYSIS - CEMENTLESS - 135° - 9 (K170910) LOT 1811227: 49 ITEMS MANUFACTURED AND RELEASED ON 21-FEB-2019. EXPIRATION DATE: 2024-02-11. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 43 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ANATOMICAL SHOULDER SYSTEM 04.01.0130 HC PEGGED GLENOID Ø44 (K170910) LOT 2103247: 26 ITEMS MANUFACTURED AND RELEASED ON 30-APR-2021. EXPIRATION DATE: 2026-04-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 16 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ANATOMICAL SHOULDER SYSTEM 04.01.0089 DOUBLE ECCENTER (K170910) LOT 2003985: 117 ITEMS MANUFACTURED AND RELEASED ON 19-JAN-2021. EXPIRATION DATE: 2025-12-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 105 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT 1 YEAR AND 1 MONTH FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO NOT BEING ABLE TO HAVE FULL ARM ROTATION. RANGE OF MOTION WAS LIMITED AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED ALL COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105167 ANATOMICAL SHOULDER SYSTEM METAL HUMERAL HEAD Ø46 SHOULDER METAL HEAD KWS MEDACTA INTERNATIONAL SA 1903164 07630040713160

Patients

Seq Age Sex Outcome Treatment
1 34 YR Male Required Intervention