FDA Adverse Event Malfunction Summary report: N

ALARIS SYSTEM

MDR report key: 17866738 · Received October 4, 2023

Report

Report Number
2016493-2023-235131
Event Type
Malfunction
Date Received
October 4, 2023
Date of Event
September 7, 2023
Report Date
September 26, 2023
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801549
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BD TECHNICAL SUPPORT TROUBLESHOOT WITH CUSTOMER OVER THE PHONE. A DEVICE HISTORY RECORD REVIEW IS PERFORMED ON EACH DEVICE REPORTED IN A MDR REPORTABLE EVENT ALONG WITH OTHER METHODS OF INVESTIGATION AS CODED IN SECTION H6 OF THIS MDR REPORT. PER 803.52(F)(11)(III) THE INFORMATION PROVIDED REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. THE COMPLAINANT OR REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO THE MANUFACTURER. H3 OTHER TEXT : DEVICE WAS NOT RETURNED TO MANUFACTURING FACILITY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PCUS WITH NEWER BATTERIES WERE UNABLE TO "PASS" THE BATTERY CONDITIONING TEST (RESULTS RANGING FROM 3801 MAH TO 3942 MAH). THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE OCCURRED DURING DEVICE MAINTENANCE. BD TECHNICAL SUPPORT REACHED OUT TO THE CUSTOMER AND PROVIDED THE FOLLOWING GUIDANCE: A COPY OF SERVICE BULLETIN 638A RELEASE OF BD ALARIS¿ SYSTEM V12.1.2 AND SERVICE BULLETIN 592A UPDATED BATTERY CARE AND BATTERY CONDITIONING TEST WAS PROVIDED TO THE CUSTOMER. ADDITIONALLY, BD TECHNICAL SUPPORT INFORMED THE CUSTOMER THAT "PER THE SERVICE BULLETIN 592A UPDATED BATTERY CARE AND BATTERY CONDITIONING TEST, IF THE NEW BATTERY CAPACITY VALUE IS BETWEEN 4000 MAMPHR (MAH) AND 4500 MAMPHR (MAH), THEN THE BATTERY CAPACITY IS SUFFICIENT FOR USE. HOWEVER, THIS IS AN OVERESTIMATE OF THE ACTUAL BATTERY CAPACITY. THIS APPLIES TO NEWLY INSTALLED BATTERIES. THEN THE UNIT WOULD NEED TO HAVE RESET THE BATTERY CAPACITANCE. THIS ONLY APPLIES TO UNITS THAT HAVE POINT OF CARE VERSION 12.1.0 AND EARLIER. PER SERVICE BULLETIN 638A RELEASE OF BD ALARIS¿ SYSTEM V12.1.2, THE BATTERY OVER ESTIMATION WAS CORRECTED. WHEN PERFORMING THE BATTERY RECONDITION WITH VERSION 12.1.2 AND ABOVE, THE ACCEPTABLE RANGE BETWEEN 3700 MAMPHR (MAH) AND 4500 MAMPHR (MAH)." SERVICE BULLETIN 592A UPDATED BATTERY CARE AND BATTERY CONDITIONING TEST ALSO NOTES THAT "IF THE NEW BATTERY CAPACITY IS BETWEEN 3700 MAMPHR (MAH) AND 3999 MAMPHR (MAH), YOU CAN PERFORM THE MANUAL BATTERY CONDITIONING TEST ON PAGE 5 TO GET A MORE ACCURATE ESTIMATE OF BATTERY CAPACITY TO DETERMINE WHETHER THE BATTERY CAN BE USED. ALTERNATIVELY, REPLACE WITH A NEW BATTERY. PROCEED TO STEP 11 TO POWER DOWN THE DEVICE AND COMPLETE THE TEST. IF THE NEW BATTERY CAPACITY IS LESS THAN 3700 MAMPHR (MAH) OR GREATER THAN 4500 MAMPHR (MAH), THIS IS OUTSIDE THE RANGE OF ACCEPTABLE BATTERY CAPACITY READING. PROCEED TO STEP 11 TO POWER DOWN THE DEVICE AND COMPLETE THE TEST. REPLACE WITH NEW BATTERY." AFTER BD TECHNICAL SUPPORT TEAM REVIEWED THE SPREADSHEET (LISTING THE BATTERY CONDITIONING TEST VALUES) PROVIDED BY THE CUSTOMER, IT WAS CONCLUDED THAT "THE UNITS ARE WORKING AS INTENDED AND ARE WITHIN THE ACCEPTABLE RANGE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1353652 ALARIS SYSTEM PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801549

Patients

Seq Age Sex Outcome Treatment
1 Unknown