FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TEST STRIPS
MDR report key: 1786449
·
Received July 27, 2010
Report
- Report Number
- 1826988-2010-00497
- Event Type
- Malfunction
- Date Received
- July 27, 2010
- Date of Event
- July 1, 2010
- Report Date
- July 10, 2010
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K062058
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
QA EVALUATED THE RETURNED REAGENT AND FOUND IT TO READ AN AVERAGE OF 405 MG/DL HIGH, OUT OF SPEC. SATISFACTORY PERFORMANCE USING IQA REAGENT.
Description of Event or Problem · 1
CUSTOMER RECEIVED A CONTROL READING OF 493 MG/DL WHICH IS HIGH OUT OF SPEC. THE NORMAL CONTROL RANGE IS 106-146 MG/DL. REPLACEMENT STRIPS WERE SENT TO THE CUSTOMER. CUSTOMER'S STRIPS HAVE BEEN EVALUATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7098C | 9HC3C03A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |