FDA Adverse Event Malfunction Summary report: N

WORKMATE CPU

MDR report key: 1785962 · Received July 30, 2010

Report

Report Number
2184149-2010-00012
Event Type
Malfunction
Date Received
July 30, 2010
Date of Event
January 6, 2009
Report Date
January 6, 2009
Manufacturer
ST. JUDE MEDICAL, AF DIVISION
Product Code
DQK
PMA / PMN Number
K092810
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FUNCTIONAL TESTING OF THE RETURNED UNIT PRODUCED AN ERROR MESSAGE WHICH CONFIRMED THE CUSTOMER COMPLAINT. THE CUSTOMER WAS ISSUED A REPLACEMENT COMPUTER.

Description of Event or Problem · 1

IT WAS REPORTED WHILE PACING DURING A PROCEDURE, A BEEP WAS HEARD AND THE EP WORKMATE SCREENS WENT BLANK. THE EP4 TOUCH SCREEN REMAINED ON. THE USER WAS UNABLE TO STOP PACING ON THE EP WORKMATE KEYBOARD AND UNABLE TO STOP PACING ON THE EP4 TOUCH SCREEN. THE SYSTEM WAS RESTARTED AND THE USER HIT BEGIN/REVIEW STUDY AND WAS THEN ABLE TO STOP PACING VIA THE KEYBOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WORKMATE CPU EP WORKMATE DQK ST. JUDE MEDICAL, AF DIVISION WM-CPU-WIN UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK