FDA Adverse Event
Malfunction
Summary report: N
WORKMATE CPU
MDR report key: 1785962
·
Received July 30, 2010
Report
- Report Number
- 2184149-2010-00012
- Event Type
- Malfunction
- Date Received
- July 30, 2010
- Date of Event
- January 6, 2009
- Report Date
- January 6, 2009
- Manufacturer
- ST. JUDE MEDICAL, AF DIVISION
- Product Code
- DQK
- PMA / PMN Number
- K092810
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FUNCTIONAL TESTING OF THE RETURNED UNIT PRODUCED AN ERROR MESSAGE WHICH CONFIRMED THE CUSTOMER COMPLAINT. THE CUSTOMER WAS ISSUED A REPLACEMENT COMPUTER.
Description of Event or Problem · 1
IT WAS REPORTED WHILE PACING DURING A PROCEDURE, A BEEP WAS HEARD AND THE EP WORKMATE SCREENS WENT BLANK. THE EP4 TOUCH SCREEN REMAINED ON. THE USER WAS UNABLE TO STOP PACING ON THE EP WORKMATE KEYBOARD AND UNABLE TO STOP PACING ON THE EP4 TOUCH SCREEN. THE SYSTEM WAS RESTARTED AND THE USER HIT BEGIN/REVIEW STUDY AND WAS THEN ABLE TO STOP PACING VIA THE KEYBOARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WORKMATE CPU | EP WORKMATE | DQK | ST. JUDE MEDICAL, AF DIVISION | WM-CPU-WIN | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |