FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE¿ NEEDLE

MDR report key: 17857235 · Received October 2, 2023

Report

Report Number
1213809-2023-01092
Event Type
Malfunction
Date Received
October 2, 2023
Date of Event
September 18, 2023
Report Date
October 24, 2023
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
10612479262598
PMA / PMN Number
K951254
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH ADDITIONAL LOT NUMBER IS AS FOLLOWS: D.4. MEDICAL DEVICE LOT #: 2202913, D.4. MEDICAL DEVICE EXPIRATION DATE: 30-JUN-2023, H.4. DEVICE MANUFACTURE DATE: 21-JUL-2022, D.4. MEDICAL DEVICE LOT #: 2202914, D.4. MEDICAL DEVICE EXPIRATION DATE: 30-JUN-2023, H.4. DEVICE MANUFACTURE DATE: 21-JUL-2022. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

ROW 1: 52 SAMPLES WERE RECEIVED BY BD. 51 CAME IN SEALED PACKAGING BLISTERS AND ONE SAMPLE WITH NO PACKAGING BLISTER. A QUALITY ENGINEER WAS ABLE TO REVIEW THE SAMPLES FROM LOT #2202913 REGARDING ITEM #306616. VISUAL INSPECTION WAS PERFORMED. NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. EACH SAMPLE WAS CONNECTED TO A SYRINGE WITH SALINE SOLUTION FOR A FUNCTIONAL TEST. THEY ALL EXPELLED THE SOLUTION WITH NORMAL FLOW. BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. WE WILL CONTINUE MONITORING THE COMPLAINT TREND FOR THE PRODUCT AND SYMPTOM. PREVIOUS INVESTIGATIONS HAVE REVEALED THAT PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED BATCH NUMBER 2202913 SHOWING NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROW 2: NO SAMPLES OR PHOTOS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PREVIOUS INVESTIGATIONS HAVE REVEALED THAT PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE NARRATIVE BELOW.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SAFETYGLIDE¿ NEEDLE PLUNGER MOVEMENT DIFFICULT. WITH LOT #'S 2202913 , 2202914. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UNABLE TO PUSH ANYTHING THROUGH IT, CAN'T PUSH ANYTHING OUT OF IT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1154554 BD SAFETYGLIDE¿ NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL SYSTEMS 2202914 10612479262598

Patients

Seq Age Sex Outcome Treatment
1 Unknown