BD SAFETYGLIDE¿ NEEDLE
Report
- Report Number
- 1213809-2023-01092
- Event Type
- Malfunction
- Date Received
- October 2, 2023
- Date of Event
- September 18, 2023
- Report Date
- October 24, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMI
- UDI-DI
- 10612479262598
- PMA / PMN Number
- K951254
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH ADDITIONAL LOT NUMBER IS AS FOLLOWS: D.4. MEDICAL DEVICE LOT #: 2202913, D.4. MEDICAL DEVICE EXPIRATION DATE: 30-JUN-2023, H.4. DEVICE MANUFACTURE DATE: 21-JUL-2022, D.4. MEDICAL DEVICE LOT #: 2202914, D.4. MEDICAL DEVICE EXPIRATION DATE: 30-JUN-2023, H.4. DEVICE MANUFACTURE DATE: 21-JUL-2022. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
ROW 1: 52 SAMPLES WERE RECEIVED BY BD. 51 CAME IN SEALED PACKAGING BLISTERS AND ONE SAMPLE WITH NO PACKAGING BLISTER. A QUALITY ENGINEER WAS ABLE TO REVIEW THE SAMPLES FROM LOT #2202913 REGARDING ITEM #306616. VISUAL INSPECTION WAS PERFORMED. NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. EACH SAMPLE WAS CONNECTED TO A SYRINGE WITH SALINE SOLUTION FOR A FUNCTIONAL TEST. THEY ALL EXPELLED THE SOLUTION WITH NORMAL FLOW. BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. WE WILL CONTINUE MONITORING THE COMPLAINT TREND FOR THE PRODUCT AND SYMPTOM. PREVIOUS INVESTIGATIONS HAVE REVEALED THAT PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED BATCH NUMBER 2202913 SHOWING NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROW 2: NO SAMPLES OR PHOTOS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PREVIOUS INVESTIGATIONS HAVE REVEALED THAT PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE NARRATIVE BELOW.
IT WAS REPORTED THAT THE BD SAFETYGLIDE¿ NEEDLE PLUNGER MOVEMENT DIFFICULT. WITH LOT #'S 2202913 , 2202914. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UNABLE TO PUSH ANYTHING THROUGH IT, CAN'T PUSH ANYTHING OUT OF IT.
NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1154554 | BD SAFETYGLIDE¿ NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON DICKINSON MEDICAL SYSTEMS | 2202914 | 10612479262598 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |