FDA Adverse Event Injury Summary report: N

PRECISION MONTAGE MRI

MDR report key: 17848478 · Received September 29, 2023

Report

Report Number
3006630150-2023-05931
Event Type
Injury
Date Received
September 29, 2023
Date of Event
September 11, 2023
Report Date
December 6, 2024
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729905943
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BLOCK B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-PADDLE LEADS-MRI UPN: M365SC8416700. MODEL: SC-8416-70. SERIAL:(B)(6). BATCH: 7070225.

Additional Manufacturer Narrative · 0

SC-1200 (SN: (B)(6)). THE RETURNED IPG WAS ANALYZED, PASSED ALL TESTS PERFORMED, AND EXHIBITED NORMAL DEVICE CHARACTERISTICS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS NOT GETTING ADEQUATE PAIN RELIEF. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED IPG AND LEAD WERE DISCARDED BY THE MEDICAL FACILITY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS NOT GETTING AN ADEQUATE PAIN RELIEF. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED IPG AND LEAD WERE DISCARDED BY THE MEDICAL FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353284 PRECISION MONTAGE MRI STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-1200 366811 08714729905943

Patients

Seq Age Sex Outcome Treatment
1 52 YR Female Required Intervention