BD QUINCKE SPINAL NEEDLES
Report
- Report Number
- 2618282-2023-00074
- Event Type
- Malfunction
- Date Received
- September 29, 2023
- Date of Event
- September 15, 2023
- Report Date
- October 10, 2023
- Manufacturer
- BD CARIBE LTD.
- Product Code
- BSP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: THREE PHOTOS RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IN THE FIRST PHOTO IT IS OBSERVED A NEEDLE ON A WHITE NAPKIN, RESIDUES OF SOME KIND OF GEL AND A SMEAR ON THE NAPKIN. IN THE SECOND PHOTO SEVERAL NEEDLES ARE SEEN WITH RESIDUE OF BLOOD AS SOME KIND OF GEL, ALSO SMALL SMEARS ARE OBSERVED, AND THE THIRD PHOTO SHOWS THE TYVEK WITH THE CATALOG¿S DATA. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 3041718, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO THE MANUFACTURING PROCESS AT THIS TIME.
IT WAS REPORTED THAT WHILE USING THE BD QUINCKE SPINAL NEEDLE HAD FOREIGN MATTER. THE FOLLOWING WAS RECEIVED FROM THE INITIAL REPORTER: TODAY WE USED BD SPINAL NEEDLE FOR THE APPLICATION OF BARRIGEL TO THE SPACE BETWEEN RECTUM AND PROSTATE. PRIMING THE NEEDLE WITH BARRIGEL WE FOUND SOME RESIDUAL METALLIC COLOURED SUBSTANCE ON GAUZE.
IT WAS REPORTED THAT WHILE USING THE BD QUINCKE SPINAL NEEDLE HAD FOREIGN MATTER. THE FOLLOWING WAS RECEIVED FROM THE INITIAL REPORTER: TODAY WE USED BD SPINAL NEEDLE FOR THE APPLICATION OF BARRIGEL TO THE SPACE BETWEEN RECTUM AND PROSTATE. PRIMING THE NEEDLE WITH BARRIGEL WE FOUND SOME RESIDUAL METALLIC COLOURED SUBSTANCE ON GAUZE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353128 | BD QUINCKE SPINAL NEEDLES | ANESTHESIA CONDUCTION NEEDLE | BSP | BD CARIBE LTD. | 3041718 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |