FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 17835742 · Received September 28, 2023

Report

Report Number
3013756811-2023-133425
Event Type
Malfunction
Date Received
September 28, 2023
Date of Event
September 5, 2023
Report Date
September 28, 2023
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613762
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING INSULIN DELIVERY. CUSTOMER LOADED A NEW CARTRIDGE TO RESOLVE THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 132-135 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2095514 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00850006613762

Patients

Seq Age Sex Outcome Treatment
1 68 YR Male INSULIN: HUMALOG