BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2023-00197
- Event Type
- Malfunction
- Date Received
- September 28, 2023
- Date of Event
- September 2, 2023
- Report Date
- October 10, 2023
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420215
- PMA / PMN Number
- K123903
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION SUMMARY: CATALOG: 442021, BATCH NO.: 3102598. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. AS PER PRODUCT INSERT A GRAM-STAINED SMEAR FROM CULTURE MEDIUM MAY CONTAIN SMALL NUMBERS OF NONVIABLE ORGANISMS DERIVED FROM MEDIA CONSTITUENTS, STAINING REAGENTS, IMMERSION OIL, GLASS SLIDES, AND SPECIMENS USED FOR INOCULATION. DUE TO THE NATURE OF BIOLOGICAL MATERIALS IN MEDIA PRODUCTS AND INHERENT ORGANISM VARIABILITY, THE USER SHOULD BE COGNIZANT OF POTENTIAL VARIABLE RESULTS IN THE RECOVERY OF CERTAIN MICROORGANISMS. MOLECULAR TESTS PERFORMED ON POSITIVE BLOOD CULTURES WILL DETECT BOTH VIABLE AND NON-VIABLE ORGANISMS COMMONLY FOUND IN CULTURE MEDIA. THEREFORE, MOLECULAR TEST RESULTS SHOULD BE EVALUATED IN CONJUNCTION WITH GRAM STAIN RESULTS IN ACCORDANCE WITH STANDARD-OF-CARE PRACTICES AS WELL AS MANUFACTURER¿S INSTRUCTIONS FOR USE. REFER TO CUSTOMER LETTER. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.
E.2 INITIAL REPORTER E-MAIL: (B)(6). H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) THAT THERE WAS MOLECULAR FALSE POSITIVES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HAZARD, INJURY OR ERRONEOUS RESULTS? YES. HAZARD, INJURY OR ERRONEOUS RESULTS DETAILS DID ERRONEOUS RESULTS OCCUR? YES IF YES¿DETAILED ERRONEOUS RESULTS (INCLUDE # OF ERRORS AND TYPE): MOLECULAR FP FOR C. TROPICALIS. 1. WHAT RESULT DID THE CUSTOMER OBTAIN FROM THE BD PRODUCT? MOLECULAR FP FOR C. TROPICAL, E. COLI, AND KLEB PNEUMO. 2. WHAT RESULT WAS THE CUSTOMER EXPECTING TO OBTAIN (IF DIFFERENT FROM THE OBTAINED RESULT) S. AUREUS. CUSTOMER RECEIVED MOLECULAR FP FOR C. TROPICALIS.
IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) THAT THERE WAS MOLECULAR FALSE POSITIVES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HAZARD, INJURY OR ERRONEOUS RESULTS? YES. HAZARD, INJURY OR ERRONEOUS RESULTS DETAILS DID ERRONEOUS RESULTS OCCUR? YES. IF YES¿DETAILED ERRONEOUS RESULTS (INCLUDE # OF ERRORS AND TYPE): MOLECULAR FP FOR C. TROPICALIS. 1. WHAT RESULT DID THE CUSTOMER OBTAIN FROM THE BD PRODUCT? MOLECULAR FP FOR C. TROPICAL, E. COLI, AND KLEB PNEUMO. 2. WHAT RESULT WAS THE CUSTOMER EXPECTING TO OBTAIN (IF DIFFERENT FROM THE OBTAINED RESULT) S. AUREUS. CUSTOMER RECEIVED MOLECULAR FP FOR C. TROPICALIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589920 | BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) | SYSTEM, BLOOD CULTURING | MDB | BECTON DICKINSON CARIBE LTD. | 3102598 | 00382904420215 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |