FDA Adverse Event
Malfunction
Summary report: N
HF CONNECTION CABLE MONO L 3M
MDR report key: 17828134
·
Received September 27, 2023
Report
- Report Number
- 9611102-2023-00056
- Event Type
- Malfunction
- Date Received
- September 27, 2023
- Date of Event
- June 16, 2023
- Report Date
- September 27, 2023
- Manufacturer
- RICHARD WOLF GMBH
- Product Code
- GEI
- UDI-DI
- 04055207000694
- PMA / PMN Number
- K982677
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 0
THE USER FACILITY HAS INFORMED RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) OF AN ISSUE REGARDING AN HF CONNECTION CABLE MONO L 3M, PART ID: 8106.033, BATCH # 747/675. ACCORDING TO THE RECEIVED INFORMATION, DURING A SCHEDULED LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, A RN CIRCULATOR WAS ATTEMPTING TO PLUG IN BOVIE CORD WHEN A SPARK FROM THE CORD OCCURRED. CORD INSPECTION REVEALED AN INSULATION FRAY WHERE THE SPARK HAPPENED. THERE WAS A REPORTED 5-10 MINUTES DELAY IN THE PROCEDURE. HOWEVER, CORD WAS REMOVED AND REPLACED WITHOUT FURTHER INCIDENT. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1019350 | HF CONNECTION CABLE MONO L 3M | GEI | RICHARD WOLF GMBH | 8106033 | 747/675 | 04055207000694 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Female |