FDA Adverse Event Malfunction Summary report: N

HF CONNECTION CABLE MONO L 3M

MDR report key: 17828134 · Received September 27, 2023

Report

Report Number
9611102-2023-00056
Event Type
Malfunction
Date Received
September 27, 2023
Date of Event
June 16, 2023
Report Date
September 27, 2023
Manufacturer
RICHARD WOLF GMBH
Product Code
GEI
UDI-DI
04055207000694
PMA / PMN Number
K982677
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

THE USER FACILITY HAS INFORMED RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) OF AN ISSUE REGARDING AN HF CONNECTION CABLE MONO L 3M, PART ID: 8106.033, BATCH # 747/675. ACCORDING TO THE RECEIVED INFORMATION, DURING A SCHEDULED LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, A RN CIRCULATOR WAS ATTEMPTING TO PLUG IN BOVIE CORD WHEN A SPARK FROM THE CORD OCCURRED. CORD INSPECTION REVEALED AN INSULATION FRAY WHERE THE SPARK HAPPENED. THERE WAS A REPORTED 5-10 MINUTES DELAY IN THE PROCEDURE. HOWEVER, CORD WAS REMOVED AND REPLACED WITHOUT FURTHER INCIDENT. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1019350 HF CONNECTION CABLE MONO L 3M GEI RICHARD WOLF GMBH 8106033 747/675 04055207000694

Patients

Seq Age Sex Outcome Treatment
1 66 YR Female