FDA Adverse Event
Malfunction
Summary report: N
T:SLIM G5 SYSTEM
MDR report key: 17828085
·
Received September 27, 2023
Report
- Report Number
- 3013756811-2023-137741
- Event Type
- Malfunction
- Date Received
- September 27, 2023
- Date of Event
- September 4, 2023
- Report Date
- September 27, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OYC
- UDI-DI
- 00853052007318
- PMA / PMN Number
- P140015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. USE ERROR. PER THE USER GUIDE: WARNING - NEVER FILL YOUR TUBING WHILE YOUR INFUSION SET IS CONNECTED TO YOUR BODY. ALWAYS ENSURE THAT THE INFUSION SET IS DISCONNECTED FROM YOUR BODY BEFORE FILLING THE TUBING. FAILURE TO DISCONNECT YOUR INFUSION SET FROM YOUR BODY BEFORE FILLING THE TUBING CAN RESULT IN OVER DELIVERY OF INSULIN.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE CUSTOMER INADVERTENTLY PERFORMED THE LOAD SEQUENCE WHILE CONNECTED TO THE SITE. THE CUSTOMER'S BLOOD GLUCOSE WAS 142-184 MG/DL. TANDEM TECHNICAL SUPPORT EDUCATED THE CUSTOMER TO NOT BE CONNECTED TO THE PUMP DURING THE FILL TUBING PROCESS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 830916 | T:SLIM G5 SYSTEM | CONTINUOUS GLUCOSE MONITOR | OYC | TANDEM DIABETES CARE | 1000354 | 00853052007318 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Female |