FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 17827718 · Received September 27, 2023

Report

Report Number
9617229-2023-15536
Event Type
Injury
Date Received
September 27, 2023
Date of Event
September 5, 2023
Report Date
November 20, 2023
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION AND OTHER-TECHNICAL WAS RECEIVED ON OCT 17, 2023. WITH LOT NUMBER#: 1862127. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: DEFLATION: OBSERVED, ONE OPENING ON THE DIAPHRAGM VALVE SIDE ASSESSED, AS CRACKED VALVE SEAT DIAPHRAGM VALVE. ONE OPENING ON THE RADIUS SIDE ASSESSED, AS SURGICAL DAMAGE. AND PARTIAL DELAMINATION ON THE VALVE SIDE ASSESSED, AS ADHESIVE FAILURE. OTHER -TECHNICAL: NO OBSERVED, PARTICLES ON THE VALVE. ADDITIONAL OBSERVATIONS: CREASE OBSERVED, ON THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED, AS NO ISSUES WITH THE MANUFACTURING PROCESS ARE OBSERVED.

Additional Manufacturer Narrative · 0

CONTINUED H6 HEALTH EFFECT IMPACT CODE: F2203- IMAGING REQUIRED. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN INITIATED. IF ANY NEW, CHANGED OR CORRECTED INFORMATION IS NOTED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 0

PHYSICIAN REPORTED, A RIGHT SIDE DEFLATION. AND PARTICLES IN VALVE. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

PHYSICIAN REPORTED A RIGHT SIDE DEFLATION AND¿ PARTICLES IN VALVE ¿. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
776468 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1862127

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female Required Intervention