FDA Adverse Event Malfunction Summary report: N

JAMSHIDI BONE MARROW TRAY 11G X4 15G ASP

MDR report key: 17826114 · Received September 27, 2023

Report

Report Number
1625685-2023-00095
Event Type
Malfunction
Date Received
September 27, 2023
Date of Event
September 11, 2023
Report Date
January 9, 2024
Manufacturer
CAREFUSION, INC
Product Code
LWE
UDI-DI
10885403057236
PMA / PMN Number
K813338
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW-UP EMDR FOR DEVICE EVALUATION: ONE TRAY SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT WAS OBSERVED THAT THE PRESENCE OF A HAIR WITHIN THE TRAY; THEREFORE, THE REPORTED FAILURE MODE WAS CONFIRMED. A REVIEW OF THE INTERNAL MANUFACTURING DEVICE RECORDS AND RAW MATERIAL HISTORY FILES FOR THE REPORTED LOT NUMBER 0001511116 WAS PERFORMED AND NO RECORDED QUALITY PROBLEMS OR REJECTIONS RELATED TO THIS INCIDENT WERE FOUND. PRODUCT UNDERGOES INSPECTIONS DURING MANUFACTURING, NO ISSUES RELATED TO THE REPORTED INCIDENT WERE IDENTIFIED, ALL PROCEDURAL AND FUNCTIONAL REQUIREMENTS FOR PRODUCT RELEASE HAVE BEEN MET. THE STERILITY RECORD WAS REVIEWED, AND NO DEVIATIONS OCCURRED, STERILITY REQUIREMENTS WERE MET, AND THE PRODUCT WAS RELEASED WITHOUT ISSUE. BASED ON THE QUALITY TEAM'S INVESTIGATION, IT WAS IDENTIFIED THAT THE PROBABLE ROOT CAUSE IS TRACED TO MANUFACTURING RELATED. MANUFACTURING PERSONNEL HAVE BEEN NOTIFIED OF THIS INCIDENT AND AWARENESS TRAINING WILL BE PROVIDED. IN ADDITION, THE CME ROOM TEST RECORDS FOR VIABLE AIR TESTING, PARTICULATE AMBIENT AIR TESTING, AND QUARTERLY VIABLE SURFACE TESTING WERE REVIEWED, AND RESULTS FOR THE CME WERE BELOW THE ACTION LIMITS, NO NEGATIVE TRENDS EXIST FOR THIS TIMEFRAME. A CID HAS BEEN INITIATED. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND DEFECT WILL CONTINUE TO BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

(B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. DEVICE PROBLEM CODE: A1801. PATIENT PROBLEM CODE: F26.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT THEY FOUND A LONG BROWN HUMAN HAIR. VERBATIM: RCC RECEIVED THE COMPLAINT VIA EMAIL. EMAIL(S) AS ATTACHED. DR. GARCIA JUST HANDED ME A BONE MARROW KIT , HE WAS OPENING TO START A STERILE PROCEDURE ON A PATIENT . IN THE KIT, HE FOUND A LONG BROWN HUMAN HAIR . THIS IS QUITE CONCERNING AS THE KITS ARE SUPPOSED TO BE STERILE . ¿ TRAYS, BONE MARROW: JAMSHIDI BONE MARROW TRAY WITH DISPOSABLE 4" NEEDLE (B)(6) ¿ LAST TIME WE PURCHASED WAS ON 8/31/2023 PO # (B)(4) AT THIS TIME, WE WOULD LIKE TO BE SURE OF THE FOLLOWING. 1. THE KITS ARE PREPARED UNDER STERILE PROCESS. 2. WHERE ARE THESE KITS BEING PREPARED FROM? 3. IS THERE A STANDARD POLICY THEY FOLLOW TO PREPARE THESE KITS ? 4. THIS KIT WILL BE TAKEN BACK, AND A CREDIT FOR THE 1 KIT WILL BE APPLIED PLEASE. AS WE PERFORM STERILE BONE MARROW BIOPSY¿S REGULARLY, WE HAVE GREAT CONCERN MOVING FORWARD WITH THIS PRODUCT THAT IS BEING PREPARED AND SHIPPED IS ACTUALLY UNDER STERILE PROCESS. IF YOU CAN PLEASE CONTACT THE MEDLINE TEAM TO IDENTIFY WHAT WILL BE DONE AND HOW THE KITS ARE BEING PREPARED , BY WHOM AND ARE THEY UNDER STERILE PROCESSES ? RESPONSE RECEIVED ON 02-OCT-2023: I JUST RECEIVED THIS REQUEST FOR ACKNOWLEDGEMENT, I WILL PRINT THE SHIPPING LABEL AND SEND TO YOU ONCE I CAN PACK IT UP AND SCHEDULE RETURN. THE HAIR WAS IDENTIFIED WHEN OPENING THE STERILE TRAIN UP, THIS DID NOT AFFECT ANY PATIENT, IT WAS IDENTIFIED BEFORE THE PROCEDURE AND THEREFORE NOT UTILIZED. A NEW TRAY WAS THEN OPENED AND UTILIZED. NO ADVERSE EVENT TAKE OF OCCURRENCE WAS (B)(6) 2023.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT THEY FOUND A LONG BROWN HUMAN HAIR. VERBATIM: RCC RECEIVED THE COMPLAINT VIA EMAIL. EMAIL(S) AS ATTACHED. DR. (B)(6) JUST HANDED ME A BONE MARROW KIT , HE WAS OPENING TO START A STERILE PROCEDURE ON A PATIENT . IN THE KIT, HE FOUND A LONG BROWN HUMAN HAIR . THIS IS QUITE CONCERNING AS THE KITS ARE SUPPOSED TO BE STERILE . ¿ TRAYS, BONE MARROW: JAMSHIDI BONE MARROW TRAY WITH DISPOSABLE 4" NEEDLE BXTBAK4511. ¿ LAST TIME WE PURCHASED WAS ON 8/31/2023 PO # (B)(6). AT THIS TIME, WE WOULD LIKE TO BE SURE OF THE FOLLOWING. 1. THE KITS ARE PREPARED UNDER STERILE PROCESS. 2. WHERE ARE THESE KITS BEING PREPARED FROM? 3. IS THERE A STANDARD POLICY THEY FOLLOW TO PREPARE THESE KITS ? 4. THIS KIT WILL BE TAKEN BACK, AND A CREDIT FOR THE 1 KIT WILL BE APPLIED PLEASE. AS WE PERFORM STERILE BONE MARROW BIOPSY¿S REGULARLY, WE HAVE GREAT CONCERN MOVING FORWARD WITH THIS PRODUCT THAT IS BEING PREPARED AND SHIPPED IS ACTUALLY UNDER STERILE PROCESS. IF YOU CAN PLEASE CONTACT THE MEDLINE TEAM TO IDENTIFY WHAT WILL BE DONE AND HOW THE KITS ARE BEING PREPARED , BY WHOM AND ARE THEY UNDER STERILE PROCESSES?

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
974011 JAMSHIDI BONE MARROW TRAY 11G X4 15G ASP BONE MARROW COLLECTION/TRANSFUSION KIT LWE CAREFUSION, INC 0001511116 10885403057236

Patients

Seq Age Sex Outcome Treatment
1 Unknown