FDA Adverse Event Malfunction Summary report: N

COBAS® SARS-COV-2 & INFLUENZA A/B

MDR report key: 17823558 · Received September 27, 2023

Report

Report Number
2243471-2023-03116
Event Type
Malfunction
Date Received
September 27, 2023
Date of Event
August 31, 2023
Report Date
September 27, 2023
Manufacturer
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
Product Code
QLT
PMA / PMN Number
EUA201779
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE OBSERVED DISCREPANCY IS MOST LIKELY DUE TO THE SAMPLE BEING A WEAK POSITIVE FOR THE FLU A THAT IS AT/NEAR THE LIMIT OF DETECTION (LOD) OF THE ASSAY. LOW VIRAL LOAD SPECIMENS THAT ARE NEAR THE ASSAY LOD MAY NOT GENERATE CONSISTENT RESULTS UPON REPEAT TESTING ACCORDING TO EXPECTED STATISTICAL VARIANCES IN DETECTION. THIS IS A SAMPLE-SPECIFIC ISSUE AND THE REAGENT IS PERFORMING AS INTENDED.

Description of Event or Problem · 0

IN LIGHT OF THE COVID-19 PANDEMIC AND THE SUBSEQUENT EMERGENCY USE AUTHORIZATIONS (EUAS) FOR SARS-COV-2 DIAGNOSTIC TESTS, THE AGENCY HAS REQUESTED HEIGHTENED REPORTING BEYOND THE REASONABLY SUGGESTS REQUIREMENTS OF 803 TO INCLUDE ALLEGATIONS OF FALSE POSITIVE OR FALSE NEGATIVE RESULTS INDEPENDENT OF HARM OR MALFUNCTION OR OFF-LABEL USE. PURSUANT TO THE AGENCY¿S INSTRUCTION, WE HEREBY SUBMIT THIS MDR. A CUSTOMER FROM THE UNITED KINGDOM ALLEGED A DISCREPANT RESULT FOR A SINGLE PATIENT WHILE USING THE COBAS® SARS-COV-2 & INFLUENZA A/B NUCLEIC ACID TEST ON THE COBAS® LIAT® SYSTEM. THE ALLEGED SAMPLE INITIALLY GENERATED A POSITIVE RESULT FOR THE SARS-COV-2 TARGET AND NEGATIVE RESULTS FOR THE INFLUENZA A AND INFLUENZA B TARGETS. THE SAME SAMPLE WAS RETESTED TWICE ON THE COBAS LIAT ANALYZER AND GENERATED NEGATIVE RESULTS FOR ALL TARGETS. A RECOLLECTED SAMPLE WAS TESTED ON A COMPETITOR ASSAY (GENEXPERT) AND GENERATED A NEGATIVE RESULT FOR SARS-COV-2. NO HARM WAS ALLEGED. AN INVESTIGATION WAS CONDUCTED TO EVALUATE THE CUSTOMER ISSUE. PER FDA¿S EUA GUIDANCE, 1 MDR WILL BE FILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
984693 COBAS® SARS-COV-2 & INFLUENZA A/B COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES QLT ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG 21107Z

Patients

Seq Age Sex Outcome Treatment
1 Unknown