FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1782081
·
Received February 11, 2008
Report
- Report Number
- 1527736-2008-00774
- Event Type
- Malfunction
- Date Received
- February 11, 2008
- Report Date
- January 9, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY. THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AN NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
THE DEVICE WAS RECEIVED WITH NO INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC SCALPEL HAND PIECE | LFL | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |