LIGACLIP MCA LARGE APPLIER
Report
- Report Number
- 1527736-2008-00576
- Event Type
- Malfunction
- Date Received
- January 30, 2008
- Date of Event
- September 8, 2007
- Report Date
- October 19, 2007
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- GDO
- PMA / PMN Number
- K820837
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE ANALYSIS RESULTS FOR THE MCL20 INSTRUMENT CONFIRMED THAT IT WAS RETURNED NON-FUNCTIONAL. THE INSTRUMENT WAS CYCLED AND WOULD NOT FEED THE CLIPS AND THE ANTI-BACKUP WAS NOTED TO BE NON-FUNCTIONAL. UPON DISASSEMBLY OF THE INSTRUMENT THE FEEDBAR WAS FOUND TO BE BENT AND DISENGAGED FROM THE FEEDBAR DRIVER THEREFORE NOT ALLOWING THE PROPER FEEDING OF THE CLIPS INTO THE JAWS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE FOUND DAMAGE. ADDITIONALLY THE CARTRIDGE COVER WAS NOTED TO BE CRACKED. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT CLIPS WOULD NOT ADVANCE. PROCEDURE: PROSTATECTOMY. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP MCA LARGE APPLIER | GDO | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | V4H82T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |