FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE
MDR report key: 1781771
·
Received June 25, 2007
Report
- Report Number
- 2954323-2007-11149
- Event Type
- Malfunction
- Date Received
- June 25, 2007
- Date of Event
- May 27, 2007
- Report Date
- June 25, 2007
- Manufacturer
- ABBOTT DIABETES CARE INC, USA
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NO INFORMATION
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES OR ERRORS WERE OBSERVED; ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION DURING CONTROL SOLUTION TESTING. THE METER WAS RETURNED AND ACCORDING TO THE MEMORY LOG, THE REPORTED READINGS COULD NOT BE FOUND.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR FREESTYLE BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 468 MG/DL AND 129 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC, USA | NA | 0631440 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |