FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 17815571 · Received September 26, 2023

Report

Report Number
3001421318-2023-10762
Event Type
Malfunction
Date Received
September 26, 2023
Date of Event
September 14, 2022
Report Date
September 25, 2023
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
00730002856789
PMA / PMN Number
K201306
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 15 MONTHS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS BEING PREPARED FOR TREATMENT. THE ROOT-CAUSE FOR THE EXPIRATORY VALVE ISSUE HAS BEEN A DEFECTIVE EXPIRATORY VALVE. THE PROBLEM WAS CORRECTED BY REPLACING THE DEFECTIVE EXPIRATORY VALVE. THERE WAS NO REPORTED HARM TO PATIENT, USER OR THIRD PARTY.

Description of Event or Problem · 0

SELF TEST FAILED TC 231032, 231001.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723633 HAMILTON MEDICAL AG HAMILTON-C3 CBK HAMILTON MEDICAL AG HAMILTON-C3 00730002856789

Patients

Seq Age Sex Outcome Treatment
1 Unknown