FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 17814840 · Received September 25, 2023

Report

Report Number
3004753838-2023-194539
Event Type
Malfunction
Date Received
September 25, 2023
Date of Event
September 2, 2023
Report Date
January 18, 2024
Manufacturer
DEXCOM, INC.
Product Code
QBJ
UDI-DI
00386270001627
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED. A REVIEW OF A BIN FILE WAS PERFORMED BY FFA AND TRANSMITTER FAILED ERROR WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723587 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9445-24 5325655 00386270001627

Patients

Seq Age Sex Outcome Treatment
1 15 YR Male