FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM

MDR report key: 1781365 · Received August 3, 2010

Report

Report Number
2134265-2010-03286
Event Type
Malfunction
Date Received
August 3, 2010
Date of Event
July 6, 2010
Report Date
July 8, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE DEFLATION DIFFICULTIES AND WITHDRAWAL RESISTANCE OCCURRED. THE 90-99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED DISTAL LEFT ANTERIOR DESCENDING ARTERY (LAD). ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY AND A MAVERICK BALLOON WAS ADVANCED TO THE LESION FOR PREDILATION. A 3.0X32MM PROMUS ELEMENT STENT WAS THEN ADVANCED TO THE LAD AND WAS SUCCESSFULLY DEPLOYED AT 16ATMS. WHEN THE PHYSICIAN ATTEMPTED TO RETRIEVE THE STENT DELIVERY BALLOON FROM THE DEPLOYED STENT HE EXPERIENCED 'FRICTION' AND IT WAS ALSO NOTED THAT THE BALLOON WAS NOT FULLY DEFLATED. THE STENT DELIVERY SYSTEM (SDS) WAS ABLE TO BE REMOVED FROM THE PATIENT AND THE DEPLOYED PROMUS ELEMENT STENT STAYED WELL POSITIONED AND APPOSED. TWO ADDITIONAL UNSPECIFIED STENTS WERE IMPLANTED IN AN UNSPECIFIED LESION AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED WITH THE PATIENT'S CURRENT CONDITION LISTED AS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911332300 13368275

Patients

Seq Age Sex Outcome Treatment
1 54 YR LAUNCHER EBU 3.5 GUIDE CATHETER| MAVERICK BALLOON CATHETER| BMW GUIDE WIRE