FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAP PRDGM INS V2.2 PL EN
MDR report key: 1781361
·
Received July 29, 2010
Report
- Report Number
- 3004209178-2010-82245
- Event Type
- Malfunction
- Date Received
- July 29, 2010
- Date of Event
- June 28, 2010
- Report Date
- July 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS, MED REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY FOLLOWED BY A CONSTANT TONE AS A RESULT OF MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY. IN ADDITION, MOISTURE DAMAGE WAS FOUND ON THE MOTOR AND BATTERY ASSEMBLY. FURTHER TESTING WAS NOT PERFORMED DUE TO THE FROZEN DISPLAY AND CONSTANT TONE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY. TROUBLESHOOTING WAS PERFORMED AND THE BUTTONS WERE UNRESPONSIVE. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAP PRDGM INS V2.2 PL EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS, MED REL | MMT-722NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |