FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-522NAL PRDGM INS V2.2 CL EN
MDR report key: 1781352
·
Received July 29, 2010
Report
- Report Number
- 3004209178-2010-82236
- Event Type
- Malfunction
- Date Received
- July 29, 2010
- Date of Event
- June 26, 2010
- Report Date
- July 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY. THE DEVICE ALSO HAD A CRACKED CASE AT THE DISPLAY WINDOW.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT SHE WENT INTO THE POOL WHILE WEARING THE INSULIN PUMP. TROUBLESHOOTING WAS PERFORMED AND MOISTURE IN THE DEVICE WAS FOUND. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-522NAL PRDGM INS V2.2 CL EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED REL | MMT-522NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |