FDA Adverse Event Malfunction Summary report: N

PUMP MMT-522NAL PRDGM INS V2.2 CL EN

MDR report key: 1781352 · Received July 29, 2010

Report

Report Number
3004209178-2010-82236
Event Type
Malfunction
Date Received
July 29, 2010
Date of Event
June 26, 2010
Report Date
July 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY. THE DEVICE ALSO HAD A CRACKED CASE AT THE DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SHE WENT INTO THE POOL WHILE WEARING THE INSULIN PUMP. TROUBLESHOOTING WAS PERFORMED AND MOISTURE IN THE DEVICE WAS FOUND. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522NAL PRDGM INS V2.2 CL EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED REL MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 40 YR