FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAS PRDGM INS V2.2 SK EN
MDR report key: 1781350
·
Received July 29, 2010
Report
- Report Number
- 3004209178-2010-82234
- Event Type
- Malfunction
- Date Received
- July 29, 2010
- Date of Event
- June 26, 2010
- Report Date
- July 15, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A BLANK DISPLAY AS A RESULT OF A BROKEN SOLDER JOINT ON THE INTERFACE BOARD.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE BLOOD GLUCOSE READING AT TIME OF THE CALL WAS 357 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT HE WOKE UP AND THE DISPLAY WAS BLANK. THE BATTERY WAS CHANGED AND THE BUTTONS WERE NOT RESPONDING. THE CUSTOMER STATED THAT HE FELL INTO POOL WITH THE DEVICE ON TWO WEEKS AGO, BUT THE INSULIN PUMP WAS FUNCTIONING AT THE TIME. NO FURTHER INFO PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |