FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAP PRDGM INS V2.2 PL EN

MDR report key: 1781333 · Received July 29, 2010

Report

Report Number
3004209178-2010-82223
Event Type
Malfunction
Date Received
July 29, 2010
Date of Event
June 25, 2010
Report Date
July 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A BLANK DISPLAY AS A RESULT OF A BROKEN SOLDER JOINT ON THE INTERFACE BOARD. FURTHER TESTING WAS NOT PERFORMED DUE TO THE BLANK DISPLAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD FAILED MULTIPLE TIMES THE BATTERY TEST AND THE DEVICE HAD A BLANK DISPLAY. THE BLOOD GLUCOSE READING AT TIME OF THE CALL WAS 390MG/DL. TROUBLESHOOTING WAS PERFORMED AND THE DEVICE WAS UNRESPONSIVE. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAP PRDGM INS V2.2 PL EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 15 YR