FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3100 SYSTEM

MDR report key: 1781316 · Received July 28, 2010

Report

Report Number
3002158293-2010-00785
Event Type
Malfunction
Date Received
July 28, 2010
Date of Event
June 25, 2010
Report Date
July 27, 2010
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE DISCHARGED BATTERY PACK WAS DEFECTIVE CELLS WITHIN THE BATTERY PACK. ONE OF THE CELLS HAD RESIDUE/CORROSION ON THEM. THE SOURCE OF THE RESIDUE/CORROSION HAS NOT BEEN POSITIVELY IDENTIFIED BUT IS LIKELY ELECTROLYTE FROM A LEAKING CELL. THE ROOT CAUSE OF THE LEAKING CELL WAS NOT POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.

Description of Event or Problem · 1

THE WIFE OF A (B)(6) MALE PT CALLED IN TO ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT ONE OF THE PT'S BATTERIES IS GIVING THE "BATTERY PACK FAULT" LED WHEN ON THE CHARGER. SUPPORT ISSUED THE PT A REPLACEMENT BATTERY PACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3100 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3100 NA

Patients

Seq Age Sex Outcome Treatment
1 56 YR