FDA Adverse Event
Injury
Summary report: N
PUMP MMT-722NAP PRDGM INS V2.2 PL EN
MDR report key: 1781302
·
Received July 28, 2010
Report
- Report Number
- 3004209178-2010-82184
- Event Type
- Injury
- Date Received
- July 28, 2010
- Date of Event
- July 11, 2010
- Report Date
- July 12, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETES KETOACIDOSIS. THE BLOOD GLUCOSE READING WAS 707MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DAILY TOTALS MATCHED THE BASAL AND BOLUS. RAN A FIXED PRIME AND HIGH PRESSURE TESTS AND PASSED. IT WAS STATED THAT THE INSULIN PUMP WAS EXPOSED TO AN X-RAY MACHINE. PERFORMED A DISPLACEMENT TEST TWICE AND FAILED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAP PRDGM INS V2.2 PL EN | INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR | Hospitalization |