FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1781279 · Received July 29, 2010

Report

Report Number
3004209178-2010-82243
Event Type
Malfunction
Date Received
July 29, 2010
Date of Event
June 28, 2010
Report Date
July 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS, MED REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A BLANK DISPLAY AS A RESULT OF CORROSION AT THE LIQUID CRYSTAL DISPLAY BOARD. ALSO, THE DEVICE HAD A LOOSE DISPLAY WINDOW. FURTHER TESTING WAS NOT PERFORMED DUE TO THE BLANK DISPLAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED NO DELIVERY. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT HE REMOVED THE INFUSION SET, RAN A BOLUS AND THEN REATTACHED, BUT THE DEVICE HAD THE SAME ALARM. THE MOTHER STATED THAT THE DISPLAY OVERLAY WAS LIFTED. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS, MED REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 17 YR