FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE FREEDOM LITE
MDR report key: 1781246
·
Received August 3, 2010
Report
- Report Number
- 2954323-2010-01030
- Event Type
- Malfunction
- Date Received
- August 3, 2010
- Date of Event
- July 8, 2010
- Report Date
- July 30, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER'S METER (B)(4) WAS RETURNED AND TESTED WITH APPROVED / RETURNED TEST STRIPS LOT # 0934926. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION. ADDITIONALLY, THE REPORTED READINGS OF 370 MG/DL AND 107 MG/DL WERE FOUND IN THE DEVICE'S INTERNAL MEMORY LOG WITHIN 10 MINUTES.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 370 MG/DL AND 107 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 0934926 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |