FDA Adverse Event Malfunction Summary report: N

C-ARM

MDR report key: 1781088 · Received July 29, 2010

Report

Report Number
1720753-2010-02202
Event Type
Malfunction
Date Received
July 29, 2010
Date of Event
June 21, 2010
Report Date
July 29, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE VOLTAGE WAS ADJUSTED ON PS1. THE SYSTEM WAS THEN FOUND TO BE WORKING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM CONTINUED TO SHUT DOWN. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 C-ARM FLUOROSCOPY XRAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment