FDA Adverse Event Malfunction Summary report: N

COBAS INTEGRA 400+

MDR report key: 1780973 · Received July 26, 2007

Report

Report Number
1823260-2007-06554
Event Type
Malfunction
Date Received
July 26, 2007
Date of Event
July 6, 2007
Report Date
July 26, 2007
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K951595
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

ACCOUNT EXPERIENCED FOUR LOW PATIENT SODIUM RESULTS THAT WERE HIGHER WHEN REPEATED. THE RESULTS WERE AS FOLLOWS (INITIAL/REPEAT AT ANOTHER FACILITY/REPEAT ON THE SAME SYSTEM MMOL/L): SAMPLE 1 132/141/142. THE INCORRECT RESULTS WERE NOT REPORTED. THE FIELD SERVICE REP WAS UNABLE TO DETERMINE A CAUSE FOR THE DISCREPANT RESULTS.

Description of Event or Problem · 2

ACCOUNT EXPERIENCED FOUR LOW PATIENT SODIUM RESULTS THAT WERE HIGHER WHEN REPEATED. THE RESULTS WERE AS FOLLOWS (INITIAL/REPEAT AT ANOTHER FACILITY/REPEAT ON THE SAME SYSTEM MMOL/L): SAMPLE 2 120/129/129. THE INCORRECT RESULTS WERE NOT REPORTED. THE FIELD SERVICE REP WAS UNABLE TO DETERMINE A CAUSE FOR THE DISCREPANT RESULTS.

Description of Event or Problem · 3

ACCOUNT EXPERIENCED FOUR LOW PATIENT SODIUM RESULTS THAT WERE HIGHER WHEN REPEATED. THE RESULTS WERE AS FOLLOWS (INITIAL/REPEAT AT ANOTHER FACILITY/REPEAT ON THE SAME SYSTEM MMOL/L): SAMPLE 3 128/139/139. THE INCORRECT RESULTS WERE NOT REPORTED. THE FIELD SERVICE REP WAS UNABLE TO DETERMINE A CAUSE FOR THE DISCREPANT RESULTS.

Description of Event or Problem · 4

ACCOUNT EXPERIENCED FOUR LOW PATIENT SODIUM RESULTS THAT WERE HIGHER WHEN REPEATED. THE RESULTS WERE AS FOLLOWS (INITIAL/REPEAT AT ANOTHER FACILITY/REPEAT ON THE SAME SYSTEM MMOL/L): SAMPLE 4 128/138/139. THE INCORRECT RESULTS WERE NOT REPORTED. THE FIELD SERVICE REP WAS UNABLE TO DETERMINE A CAUSE FOR THE DISCREPANT RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS INTEGRA 400+ CLINICAL CHEMISTRY ANALYZER - JJE JJE ROCHE DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1 74 YR
2 75 YR
3 55 YR
4 53 YR