FDA Adverse Event Injury Summary report: N

UICHIN EVO HEART POSITIONER

MDR report key: 1780955 · Received July 26, 2010

Report

Report Number
2135394-2010-00011
Event Type
Injury
Date Received
July 26, 2010
Date of Event
July 8, 2010
Report Date
July 8, 2010
Manufacturer
MEDTRONIC PERFUSION SYSTEMS
Product Code
MWS
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ANALYSIS: UPON RECEIPT AT MEDTRONIC'S QUALITY LABORATORY, VISUAL INSPECTION SHOWED NO OUTWARD SIGNS OF PHYSICAL DAMAGE OR ABNORMALITIES. THERE WERE NO ROUGH SPOTS DETECTED ON THE HEADLINK OR CAPTURE FOAM. REVIEW OF MFG RECORDS FOR THIS PRODUCT DID NOT REVEAL ANY ABNORMALITIES OR NON-CONFORMANCES. CONCLUSION: THE REASON FOR RETURN COULD NOT BE CONFIRMED. ANALYSIS DID NOT REVEAL ANY AREAS ON THE URCHIN EVO THAT WOULD HAVE CAUSED DAMAGE TO THE HEART. SINCE THE DEVICE WAS TESTED AND FOUND TO BE FREE OF ABNORMALITIES AND ROUGH SPOTS, IT IS LIKELY THAT THE ROOT CAUSE OF THIS EVENT IS SIMPLY THE PT'S FRAGILE HEART TISSUE. IT WAS ALSO REPORTED THAT THE CUSTOMER RECENTLY SWITCHED TO THE MODEL HP3500 PRODUCT FROM A COMPETITOR'S. IT IS POSSIBLE THAT THE CUSTOMER IS USED TO USING A SLIDING MOTION AS SOME COMPETITIVE PRODUCTS FUNCTION. HOWEVER, WITH THE HP3500, THE DEVICE IS SUPPOSED TO "LOCK" IN PLACE. MOVING THE DEVICE IN A SLIDING MOTION AT HIGH SUCTION COULD POTENTIALLY DAMAGE A PT'S HEART TISSUE. INSTRUCTIONS FOR USE, DOCUMENT NUMBER: (B)(4), REV A DOCUMENT'S CONTRAINDICATIONS SUCH AS FRAGILE TISSUE.

Description of Event or Problem · 1

MEDTRONIC REC'D INFO THAT AFTER PLACING THIS URCHIN DEVICE TO THE APEX, THE SURGEON WAS MOVING THE HEART AND DAMAGE TO THE HEART TISSUE RESULTED. THE SURGEON COMMENTED THAT THEY FELT THE DAMAGE WAS DUE TO THE PT'S FRAGILE HEART TISSUE. THE DAMAGED TISSUE WAS ABLE TO BE REPAIRED WITHOUT FURTHER COMPLICATIONS REPORTED. THE URCHIN DEVICE WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UICHIN EVO HEART POSITIONER MWS MEDTRONIC PERFUSION SYSTEMS HP3500 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Life Threatening| R