FDA Adverse Event Malfunction Summary report: N

PERFUSOR®

MDR report key: 17807845 · Received September 25, 2023

Report

Report Number
9610825-2023-00438
Event Type
Malfunction
Date Received
September 25, 2023
Date of Event
August 17, 2023
Report Date
November 15, 2023
Manufacturer
B BRAUN MELSUNGEN AG
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4). THE DEVICE WAS NOT AVAILABLE. ONLY THE HISTORY DATA WAS SENT FOR INVESTIGATION. 2. RESULTS: 2.1 THE DEVICE HISTORY FILES WERE ANALYZED. THE INFUSION WAS STARTED ON (B)(6) 2023 AT 06:12 PM. ON (B)(6) 2023 AT 03:40 AM THE PRE ALARM "BATTERY NEAR EMPTY" OCCURRED. 30 MINUTES LATER THE ALARM "BATTERY EMPTY" OCCURRED AND THREE MINUTES LATER THE PUMP SWITCHED OFF BY ITSELF AND THE INFUSION STOPPED. NO OTHER ABNORMALITIES WERE FOUND IN THE DEVICE AND ALARM HISTORY. 3. JUDGMENT: 3.1 THE COMPLAINT COULD NOT BE CONFIRMED. BASED ON THE DEVICE AND ALARM HISTORY, NO MALFUNCTION OF THE DEVICE WAS FOUND. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN INTERNAL REPORT NUMBER (B)(4). A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN SINGAPORE: "NUH-8713030-SN(B)(6)-ALLEGED PUMPED AUTO" ACCORDING TO THE CUSTOMER: "PUMP WAS SUPPOSED TO BE RUNNING A THERAPY BUT WAS FOUND TO BE SHUT DOWN AND HAVING A VOLUME DISCREPENCY".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1810278 PERFUSOR® PUMP, INFUSION FRN B BRAUN MELSUNGEN AG 8713030

Patients

Seq Age Sex Outcome Treatment
1 Unknown