FDA Adverse Event Malfunction Summary report: N

ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT

MDR report key: 17807000 · Received September 25, 2023

Report

Report Number
3002808486-2023-00247
Event Type
Malfunction
Date Received
September 25, 2023
Report Date
October 19, 2023
Manufacturer
WILLIAM COOK EUROPE
Product Code
MIH
UDI-DI
10827002242781
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURERS REF# (B)(4). AFTER INVESTIGATION THE EVENT FOR THIS PR# IS NO LONGER REPORTABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

MANUFACTURERS REF# (B)(4). BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. G4) PMA/510(K): P180001. INVESTIGATION IS STILL IN PROGRESS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE DOCTOR STATED THAT IT WAS VERY DIFFICULT TO USE THE DEVICE DURING THE PROCEDURE. THE ENDOGRAFT WAS VERY DIFFICULT TO RELEASE. IT SEEMS LIKE IT WAS STUCK IN THE SHEATH, AS IS IT WAS A LONG STANDING DEVICE. WE HAD NO PROBLEMS OF NAVIGATION RELATED TO VESSELS TORTUOSITY. WAS IT DIFFICULT TO WITHDRAW THE SHEATH OR WAS IT DIFFICULT PULLING THE TRIGGER-WIRE TO RELEASE? BOTH MANOUVERS WERE DIFFICULT. WAS THE PHYSICIAN ABLE TO LAND THE DEVICE IN THE PLANNED TARGET ZONE. YES, WITH GREAT EFFORT AND EXPERIENCE. WAS THERE A NEED TO ROTATE THE INTRODUCTION SYSTEM DURING ADVANCEMENT. NO EVIDENT TORTUOSITY OF THE AORTA WAS DETECTED DURING THE PROCEDURE. WAS THE GREEN TRIGGER-WIRE KNOB ROTATED WHEN IT WAS WITHDRAWN? YES. PATIENT OUTCOME: A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1974659 ZENITH TX2 DISSECTION ENDOVASCULAR GRAFT STRAIGHT COMPONENT MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH WILLIAM COOK EUROPE E4357951 10827002242781

Patients

Seq Age Sex Outcome Treatment
1 Male