2010 RACK ELECSYS
Report
- Report Number
- 1823260-2007-08261
- Event Type
- Malfunction
- Date Received
- September 21, 2007
- Date of Event
- August 31, 2007
- Report Date
- September 21, 2007
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- MMI
- PMA / PMN Number
- K961481
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
TWO PT SAMPLES WITH DISCREPANT TROPONIN T RESULTS. SAMPLE FROM (B)(6) 2007, INITIAL RESULT 0.013 NG/ML, REPEAT 0.125 NG/ML. SAMPLE FROM (B)(6) 2007, INITIAL RESULT 0.01 NG/ML, REPEAT 0.125 NG/ML. A NEW SAMPLE WAS ALSO TESTED ON (B)(6) 2007 AND GAVE INITIAL RESULT OF 0.02 NG/ML, REPEAT 0.126 NG/ML. INITIAL RESULTS WERE REPORTED, PTS NOT ADVERSELY AFFECTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE PROBE WAS OUT OF ALIGNMENT AND A BELT WAS LOOSE. THE INSTRUMENT WAS REPAIRED.
TWO PT SAMPLES WITH DISCREPANT TROPONIN T RESULTS. PT 2, SAMPLE FROM (B)(6) 2007, INITIAL RESULT<0.01 NG/ML, REPEAT 0.395 NG/ML. SAME SAMPLE REPEATED ON DIFFERENT INSTRUMENT, SAME METHOD, GAVE RESULT OF 0.411 NG/ML. INITIAL RESULTS WERE REPORTED, PTS NOT ADVERSELY AFFECTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE PROBE WAS OUT OF ALIGNMENT AND A BELT WAS LOOSE. THE INSTRUMENT WAS REPAIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 2010 RACK ELECSYS | IMMUNOCHEMISTRY ANALYZER - JJE | MMI | ROCHE DIAGNOSTICS | 2010 RACK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | |||
| 2 | 85 YR |