FDA Adverse Event Malfunction Summary report: N

TRANSTAR

MDR report key: 1780094 · Received May 6, 2010

Report

Report Number
1824206-2010-06927
Event Type
Malfunction
Date Received
May 6, 2010
Date of Event
April 13, 2010
Report Date
April 13, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

TECHNICIAN FOUND STRETCHER WITH NOTE STATING BRAKE NOT WORKING. HE FOUND THE SCREW ON FOOT BRAKE/STEER CASTER LINK HAD BROKEN OFF AND LINK WAS HITTING FRAME WHEN BRAKE IS BEING APPLIED AND WOULD NOT LOCK. TECH REMOVED THE REST OF SCREW FROM HEX ROD AND REPLACED WITH NEW SCREW TO RESOLVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR WHEELED STRETCHER FPO HILL-ROM, INC. P8005A000015

Patients

Seq Age Sex Outcome Treatment
1