FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR
MDR report key: 1780094
·
Received May 6, 2010
Report
- Report Number
- 1824206-2010-06927
- Event Type
- Malfunction
- Date Received
- May 6, 2010
- Date of Event
- April 13, 2010
- Report Date
- April 13, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
TECHNICIAN FOUND STRETCHER WITH NOTE STATING BRAKE NOT WORKING. HE FOUND THE SCREW ON FOOT BRAKE/STEER CASTER LINK HAD BROKEN OFF AND LINK WAS HITTING FRAME WHEN BRAKE IS BEING APPLIED AND WOULD NOT LOCK. TECH REMOVED THE REST OF SCREW FROM HEX ROD AND REPLACED WITH NEW SCREW TO RESOLVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR | WHEELED STRETCHER | FPO | HILL-ROM, INC. | P8005A000015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |