FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 17800442 · Received September 22, 2023

Report

Report Number
1024879-2023-00637
Event Type
Malfunction
Date Received
September 22, 2023
Date of Event
August 29, 2023
Report Date
November 17, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686075
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH HAVING THEIR ECLIPSE SHIELD ACTIVATED, AND NO ISSUES WERE OBSERVED RELATING TO DEFECTIVE LOCKING MECHANISM AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THE SAFETY SHIELD BROKE OFF OF NEEDLE. THERE WAS NO REPORT OF PATIENT OR USER IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORT A NEW INCIDENT WHERE THE SAFETY SHIELD FELL OFF DURING BLOOD COLLECTION, FROM NEEDLES 368607, LOT#: 3093593.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THE SAFETY SHIELD BROKE OFF OF NEEDLE. THERE WAS NO REPORT OF PATIENT OR USER IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORT A NEW INCIDENT WHERE THE SAFETY SHIELD FELL OFF DURING BLOOD COLLECTION, FROM NEEDLES 368607, LOT#: 3093593.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1421945 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 3093593 50382903686075

Patients

Seq Age Sex Outcome Treatment
1 Unknown