FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 17794897 · Received September 21, 2023

Report

Report Number
2647876-2023-00173
Event Type
Malfunction
Date Received
September 21, 2023
Date of Event
August 28, 2023
Report Date
December 7, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: CATALOG: 442021, BATCH NO.: 3138058. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. B3: DATE OF EVENT: 28-AUG-2023. B5: DESCRIBE EVENT: REPORT 3 OF 7 IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS A MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. WHAT IS THE BIOFIRE (OR OTHER MANUFACTURER) LOT NUMBER? 3. 1556823. WHAT ORGANISMS WERE SEEN ON GRAM STAIN? 3. ENTEROCOCCUS FAECIUM, ESCHERICHIA COLI, + CANDIDA TROPICALIS- (LOT NUMBER & OCCURENCE DATE) AUGUST 28, 3138058".

Additional Manufacturer Narrative · 0

D4: MEDICAL DEVICE LOT #: UNKNOWN. D4: MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H3: A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. H4: DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 0

REPORT 3 OF 7: IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS A MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. WHAT IS THE BIOFIRE (OR OTHER MANUFACTURER) LOT NUMBER? 1. 0984323. WHAT ORGANISMS WERE SEEN ON GRAM STAIN? 1. CANDIDA GLABRATA + CANDIDA TROPICALIS (LOT NUMBER & OCCURENCE DATE) (B)(6), 3130613.

Description of Event or Problem · 0

REPORT 3 OF 7 IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS A MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. WHAT IS THE BIOFIRE (OR OTHER MANUFACTURER) LOT NUMBER? 3. 1556823. WHAT ORGANISMS WERE SEEN ON GRAM STAIN? 3. ENTEROCOCCUS FAECIUM, ESCHERICHIA COLI, + CANDIDA TROPICALIS- (LOT NUMBER & OCCURENCE DATE) AUGUST 28, 3138058".

Description of Event or Problem · 0

REPORT 3 OF 7. IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS A MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "MOLECULAR FALSE POSITIVE WITH C.TROPICALIS. WHAT IS THE BIOFIRE (OR OTHER MANUFACTURER) LOT NUMBER? 1556823. WHAT ORGANISMS WERE SEEN ON GRAM STAIN? ENTEROCOCCUS FAECALIS + CANDIDA TROPICALIS. WHAT ORGANISMS GREW ON CULTURE PLATE? ENTEROCOCCUS FAECALIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2060735 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. UNKNOWN 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 Unknown